These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.


PUBMED FOR HANDHELDS

Search MEDLINE/PubMed


  • Title: Column liquid chromatographic determination of hydrolysed bopindolol, in the picogram per millilitre range in plasma, using cartridge extraction and dual electrochemical detection.
    Author: Humbert H, Denouel J, Keller HP.
    Journal: J Chromatogr; 1987 Nov 27; 422():205-15. PubMed ID: 2893803.
    Abstract:
    A highly sensitive and specific column liquid chromatographic assay with electrochemical detection was developed for hydrolysed bopindolol, an active metabolite of bopindolol (Sandonorm) in human plasma. The pre-chromatographic sample preparation involved Extrelut column clean-up followed by liquid extraction of the organic extract into dilute acetic acid. Separation was on a Nucleosil ODS 3-microns column at 40 degrees C, with a phosphate buffer-methanol mobile phase. Detection was performed at +450 mV with an ESA electrochemical detector. Mepindolol was used as internal standard and quantitation was based on peak-area ratios. Total analysis time was 14 min per sample. The recovery rate of the assay was at least 70% for both compounds. A detection limit as low as 25 pg/ml, starting with 1 ml of plasma, was achieved. The day-to-day reproducibility and accuracy, checked with quality-control samples, demonstrated the reliability of this assay used by different analysts, on different chromatographic systems and over a long period of time.
    [Abstract] [Full Text] [Related] [New Search]