These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.


PUBMED FOR HANDHELDS

Search MEDLINE/PubMed


  • Title: Differences between values for plasma and serum in tests performed in the Ektachem 700 XR Analyzer, and evaluation of "plasma separator tubes (PST)".
    Author: Doumas BT, Hause LL, Simuncak DM, Breitenfeld D.
    Journal: Clin Chem; 1989 Jan; 35(1):151-3. PubMed ID: 2910557.
    Abstract:
    We measured 25 analytes in plasma and serum from the same blood specimen, using the Kodak Ektachem 700 XR Analyzer. For 22 of the analytes, values for plasma were practically the same as those for serum; for the other three, differences between concentrations in plasma and serum were significant, both statistically and medically. Values for the same analytes in plasma from regular heparinized tubes were essentially indistinguishable from those for plasma obtained by use of Plasma Separator Tubes (PST).
    [Abstract] [Full Text] [Related] [New Search]