  • Title: Luminol, horseradish peroxidase, and glucose oxidase ternary functionalized graphene oxide for ultrasensitive glucose sensing.
    Author: Li F, Ma W, Liu J, Wu X, Wang Y, He J.
    Journal: Anal Bioanal Chem; 2018 Jan; 410(2):543-552. PubMed ID: 29167935.
    Abstract:
    Luminol, horseradish peroxidase (HRP), and glucose oxidase (GOx) ternary functionalized graphene oxide (HRP/GOx-luminol-GO) with excellent chemiluminescence (CL) activity and specific enzymatic property was prepared via a simple and general strategy for the first time. In this approach, luminol functionalized GO (luminol-GO) was prepared by gently stirring GO with luminol. Then HRP and GOx were further co-immobilized onto the surface of luminol-GO by storing HRP and GOx with luminol-GO at 4 °C overnight, to form HRP/GOx-luminol-GO bionanocomposites. The synthesized HRP/GOx-luminol-GO could react with H2O2 generated from GOx catalyzed glucose oxidization reaction, to produce strong CL emission in the presence of co-immobilized HRP. Thus, we developed an ultrasensitive, homogeneous, reagentless, selective, and simple CL sensing system for glucose detection. The resulting biosensors exhibited ultra-wide linear range from 5.0 nM to 5.0 mM, and an ultra-low detection limit of 1.2 nM, which was more than 3 orders of magnitude lower than previously reported methods. Furthermore, the sensing system was successfully applied for the detection of glucose in human blood samples.