  • Title: Correction of motion artifacts in endoscopic optical coherence tomography and autofluorescence images based on azimuthal en face image registration.
    Author: Abouei E, Lee AMD, Pahlevaninezhad H, Hohert G, Cua M, Lane P, Lam S, MacAulay C.
    Journal: J Biomed Opt; 2018 Jan; 23(1):1-13. PubMed ID: 29302954.
    Abstract:
    We present a method for the correction of motion artifacts present in two- and three-dimensional in vivo endoscopic images produced by rotary-pullback catheters. This method can correct for cardiac/breathing-based motion artifacts and catheter-based motion artifacts such as nonuniform rotational distortion (NURD). This method assumes that en face tissue imaging contains slowly varying structures that are roughly parallel to the pullback axis. The method reduces motion artifacts using a dynamic time warping solution through a cost matrix that measures similarities between adjacent frames in en face images. We optimize and demonstrate the suitability of this method using a real and simulated NURD phantom and in vivo endoscopic pulmonary optical coherence tomography and autofluorescence images. Qualitative and quantitative evaluations of the method show an enhancement of the image quality.
    [Abstract] [Full Text] [Related] [New Search]