These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.
Pubmed for Handhelds
PUBMED FOR HANDHELDS
Search MEDLINE/PubMed
Title: Coupling imaged capillary isoelectric focusing with mass spectrometry using a nanoliter valve. Author: Montealegre C, Neusüß C. Journal: Electrophoresis; 2018 May; 39(9-10):1151-1154. PubMed ID: 29469203. Abstract: Imaged capillary isoelectric focusing (iCIEF) is a powerful separation technique applied frequently for the analysis of biotherapeutics. However, direct mass spectrometric characterization is usually not possible. Here, focused peaks from an iCIEF system are transferred to a four-port nanoliter valve where peaks of interest are cut and transferred directly (flow injection) or via capillary zone electrophoresis to electrospray mass spectrometry. At first, flow injection coupling was tested with a mixture of peptides showing intraday precision (RSD) of 20.0 and 3.4% in area and 15.2 and 6.3% in intensity for angiotensin I and leucine-enkephalin, respectively. For the analysis of charge variants in an intact mAb, flow injection and CZE as second dimension were compared, demonstrating the usability of flow injection even for large proteins. However, improved spectra quality was achieved when CZE-MS was applied. In this way, accurate masses were obtained not only for the main isoform, but also for the main acidic and basic variants. These results demonstrate the power of iCIEF-CZE-MS for the analysis of biotherapeutics.[Abstract] [Full Text] [Related] [New Search]