These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.
Pubmed for Handhelds
PUBMED FOR HANDHELDS
Search MEDLINE/PubMed
Title: Evaluation of the measurement of blood glucose levels on dried filter paper blood spots. Author: Winocour PH, McKinnon GA, McMurray JR, Anderson DC. Journal: Diabet Med; 1985 Jul; 2(4):269-71. PubMed ID: 2951081. Abstract: The measurement of blood glucose using discs punched from dried spots on filter paper has been examined as a method to monitor patients' home blood glucose results. We confirm that the method is precise, reproducible, and accurate, but show that impregnation of the filter paper with boric acid leads to an initial false elevation in measured glucose levels and does not improve stability of the sample more than does refrigeration. Degradation of blood glucose is insignificant over 28 days if samples are stored at -20 degrees C, and samples kept at 4 degrees C show minimal falls provided they are analysed within 7 days. The greatest breakdown of glucose was shown to occur during the first 48 hours after application. The filter paper glucose measurements correlated closely with parallel measurements using a Yellow Springs Glucose Analyser.[Abstract] [Full Text] [Related] [New Search]