These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.
Pubmed for Handhelds
PUBMED FOR HANDHELDS
Search MEDLINE/PubMed
Title: Non-enzymatic electrochemical platform for parathion pesticide sensing based on nanometer-sized nickel oxide modified screen-printed electrodes. Author: Khairy M, Ayoub HA, Banks CE. Journal: Food Chem; 2018 Jul 30; 255():104-111. PubMed ID: 29571455. Abstract: Nanozyme-based electrochemical sensors have attracted much attention because of their low cost, sensitivity and remarkable stability under extensive environmental and industrial conditions. Interestingly, the physical characteristics of the nanomaterials in terms of size, shape, composition, surface area and porosity dominate the electrochemical processes at electrode surfaces. Herein, we explore nickel oxide nanoplatelets (NPs) modified screen-printed electrode-based nanozyme sensors that displays high electrochemical activity including stability, sensitivity, selectivity and applicability for organophosphate pesticide (Parathion) determination. Differential pulse voltammogram of NiO-SPE in presence of parathion showed a characteristic peak current at -1.0 V (vs. Ag/AgCl). The NiO-SPE platform allows determination of parathion over the concentration range of 0.1-30 µM with a limit of detection (LOD) of 0.024 µM. The sensing platform is found to detect parathion of interferences without compromising the sensitivity of the sensor. Such interesting features offer a sensitive determination of parathion in water, urine and vegetable samples.[Abstract] [Full Text] [Related] [New Search]