These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.


PUBMED FOR HANDHELDS

Search MEDLINE/PubMed


  • Title: Large Metal Heads and Highly Cross-Linked Polyethylene Provide Low Wear and Complications at 5-13 Years.
    Author: Lachiewicz PF, O'Dell JA, Martell JM.
    Journal: J Arthroplasty; 2018 Jul; 33(7):2187-2191. PubMed ID: 29588121.
    Abstract:
    BACKGROUND: Highly cross-linked polyethylene (XLPE) is reported to have low rates of linear and volumetric wear at 10-14 years. In a prior study, larger (36 and 40 mm) femoral heads were associated with more volumetric wear, but there were only 12 hips with these heads. METHODS: We evaluated 107 hips (93 patients, with a mean age of 76 years) with one design of uncemented acetabular component, a 36 (90 hips) or 40 mm (17 hips) metal femoral head, and one electron beam 100 kGy irradiated and remelted XLPE at a mean follow-up of 8 years (range 5-13 years). Selection of these femoral heads was based on several factors, including the perceived risk of dislocation, the outer diameter size of the acetabular component, and liner availability. Measurements of linear and volumetric wear were performed in one experienced laboratory by the Martell method and analyzed using the first-to-last method. Standard radiographs, with additional Judet views, were used to detect periprosthetic osteolysis. Clinical records were used to determine the complications of dislocation, liner fracture, and painful trunnion corrosion. RESULTS: For the entire cohort, the median linear wear rate was 0.041 mm/y (95% confidence interval, 0.031-0.057) and the median volumetric wear rate was 34.6 mm3/y (95% confidence interval, 31.4-53.5). With the numbers available, there was no difference in linear or volumetric wear between the 36 and 40 mm head sizes. Small, asymptomatic osteolytic lesions were noted in 3 hips (2%). There were 3 patients (3%) with dislocation (2 early and 1 late), but these have not had a revision. There were no revisions for loosening, no liner fracture, and no patient with symptomatic trunnion corrosion. CONCLUSION: This acetabular component and XLPE with large metal heads had low rates of linear and volumetric wear. Large metal femoral heads did not lead to liner fracture, loosening, or symptomatic trunnion corrosion in this patient population. However, we recommend longer clinical follow-up studies and caution in the routine use of larger metal femoral heads in other, younger patient populations.
    [Abstract] [Full Text] [Related] [New Search]