These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.
Pubmed for Handhelds
PUBMED FOR HANDHELDS
Search MEDLINE/PubMed
Title: REVERSE PHASE HIGH PERFORMANCE LIQUID CHROMATOGRAPHY METHOD FOR DETERMINATION OF 5-FLUOROURACIL IN RABBIT PLASMA. Author: Nasir B, Muhammad Ranjha N, Hanif M, Abbas G. Journal: Acta Pol Pharm; 2017 Mar; 74(2):379-383. PubMed ID: 29624242. Abstract: A simple, efficient, accurate and selective HPLC method has been developed and validated successfully for the estimation of 5-fluorouracil in rabbit plasma. The drug was eluted by using Supelco C18 column (1.5 cm x 4.6 mm, 5 μm) with a mobile phase consisting of methanol and water (20: 80, v/v) by adjusting pH at 3.2, using perchloric acid solution. The retention time was found to be 4.107 with a flow rate of I mL/min. Multiple validation parameters evaluated with high accuracy indicating that the current method sufficiently qualifies the recommended criteria. Regression value obtained from linearity curve was R² = 0.999 and percentage recovery obtained was ranged from 96.6 to 102.5%. A fine response in short run time with perfect resolution made the method highly suitable for pharmacokinetic studies.[Abstract] [Full Text] [Related] [New Search]