  • Title: REVERSE PHASE HIGH PERFORMANCE LIQUID CHROMATOGRAPHY METHOD FOR DETERMINATION OF 5-FLUOROURACIL IN RABBIT PLASMA.
    Author: Nasir B, Muhammad Ranjha N, Hanif M, Abbas G.
    Journal: Acta Pol Pharm; 2017 Mar; 74(2):379-383. PubMed ID: 29624242.
    Abstract:
    A simple, efficient, accurate and selective HPLC method has been developed and validated successfully for the estimation of 5-fluorouracil in rabbit plasma. The drug was eluted by using Supelco C18 column (1.5 cm x 4.6 mm, 5 μm) with a mobile phase consisting of methanol and water (20: 80, v/v) by adjusting pH at 3.2, using perchloric acid solution. The retention time was found to be 4.107 with a flow rate of I mL/min. Multiple validation parameters evaluated with high accuracy indicating that the current method sufficiently qualifies the recommended criteria. Regression value obtained from linearity curve was R² = 0.999 and percentage recovery obtained was ranged from 96.6 to 102.5%. A fine response in short run time with perfect resolution made the method highly suitable for pharmacokinetic studies.
    [Abstract] [Full Text] [Related] [New Search]