These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.
Pubmed for Handhelds
PUBMED FOR HANDHELDS
Search MEDLINE/PubMed
Title: Comparative bench study evaluation of different infant interfaces for non-invasive ventilation. Author: Conti G, Spinazzola G, Gregoretti C, Ferrone G, Cortegiani A, Festa O, Piastra M, Tortorolo L, Costa R. Journal: BMC Pulm Med; 2018 Apr 07; 18(1):57. PubMed ID: 29625596. Abstract: BACKGROUND: To compare, in terms of patient-ventilator interaction and performance, a new nasal mask (Respireo, AirLiquide, FR) with the Endotracheal tube (ET) and a commonly used nasal mask (FPM, Fisher and Paykel, NZ) for delivering Pressure Support Ventilation (PSV) in an infant model of Acute Respiratory Failure (ARF). METHODS: An active test lung (ASL 5000) connected to an infant mannequin through 3 different interfaces (Respireo, ET and FPM), was ventilated with a standard ICU ventilator set in PSV. The test lung was set to simulate a 5.5 kg infant with ARF, breathing at 50 and 60 breaths/min). Non-invasive ventilation (NIV) mode was not used and the leaks were nearly zero. RESULTS: The ET showed the shortest inspiratory trigger delay and pressurization time compared to FPM and Respireo (p < 0.01). At each respiratory rate tested, the FPM showed the shortest Expiratory trigger delay compared to ET and Respireo (p < 0.01). The Respireo presented a lower value of Inspiratory pressure-time product and trigger pressure drop than ET (p < 0.01), while no significant difference was found in terms of pressure-time product at 300 and 500 ms. During all tests, compared with the FPM, ET showed a significantly higher tidal volume (VT) delivered (p < 0.01), while Respireo showed a trend toward an increase of tidal volume delivered compared with FPM. CONCLUSIONS: The ET showed a better patient-ventilator interaction and performance compared to both the nasal masks. Despite the higher internal volume, Respireo showed a trend toward an increase of the delivered tidal volume; globally, its efficiency in terms of patient-ventilator interaction was comparable to the FPM, which is the infant NIV mask characterized by the smaller internal volume among the (few) models on the market.[Abstract] [Full Text] [Related] [New Search]