These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.
Pubmed for Handhelds
PUBMED FOR HANDHELDS
Search MEDLINE/PubMed
Title: Immobilization of a phosphonium ionic liquid on a silica monolith for hydrophilic interaction chromatography. Author: Moravcová D, Planeta J, King AWT, Wiedmer SK. Journal: J Chromatogr A; 2018 Jun 01; 1552():53-59. PubMed ID: 29653778. Abstract: A methodology for preparing phosphonium-based ionic liquid modified silica-based monolithic capillary columns is presented. The silica monolithic columns with dimensions of 150 × 0.1 mm were modified by a phosphonium-based ionic liquid (trioctyl(3/4-vinylbenzyl)phosphonium chloride) via 3-(trimethoxysilyl)propyl methacrylate. The prepared columns were evaluated under hydrophilic interaction liquid chromatography separation conditions, employing a sample mixture containing purine and pyrimidine bases and nucleosides. Detection was made by UV. The high efficiency of the original silica monolith was preserved even after modification, and it reached values in the range of 98,000-174,000 theoretical plates/m. The effects of the concentration of acetonitrile in the mobile phase, the presence of additives in the mobile phase, such as, acetic acid or ammonium acetate, and the pH of the mobile phase on the separation of some selected analytes were investigated. The prepared columns showed different separation selectivity compared to silica, phenyl and sulfobetaine stationary phases.[Abstract] [Full Text] [Related] [New Search]