  • Title: [A new method for determination of cadmium in plasma (author's transl)].
    Author: Raptis S, Müller K.
    Journal: Clin Chim Acta; 1978 Sep 01; 88(2):393-402. PubMed ID: 29728.
    Abstract:
    Cadmium can be isolated from plasma by formation of negative charged iodide complexes bound to an anion exchange resin. Elution from the resin gives a cadmium solution with only a few interfering substances in an uniform matrix. Thus an undisturbed determination by flameless atomic absorption spectrometry is achieved. The mean of the natural cadmium level found in plasma was 1.9 +/- 0.8 ng/ml, within the normal range between 1.1 and 3.3 ng/ml.