These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.
Pubmed for Handhelds
PUBMED FOR HANDHELDS
Search MEDLINE/PubMed
Title: Calibration and Electrical Validation of a BIS Portable System. Author: Alves J, Sousa F, Correia MV. Journal: Stud Health Technol Inform; 2018; 249():122-129. PubMed ID: 29866967. Abstract: Heart failure (HF) affects at least 26 million people worldwide and is considered a global pandemic. Almost 1 out of 4 hospitalised patients are re-hospitalised for HF within the 30-day post-discharge period. This can be avoided if changes to the hemodynamics of the HF patient's body are early detected. Bioimpedance Spectroscopy (BIS) is a method that allows the measurement and analysis of multi-frequency body complex impedance and can be used to detect changes to the HF patient's hemodynamics. This paper presents the calibration and validation process of a low-cost portable BIS system, as well as, the study of the best Cole parameter estimation methods. The BIS system was calibrated using a three reference circuit method and a Resistance-Capacitance-Inductance (RCL) meter as calibration system. After calibration, BIS impedance measurements of validation circuits presented an average phase error of 0.76 degrees and an average magnitude relative error of 0.6% when compared with standard values. Regarding Cole parameters estimation, using the Impedance model and the Non-Linear Least Squares method for curve fitting, the relative errors were below 4% when compared with the expected values.[Abstract] [Full Text] [Related] [New Search]