  • Title: Calibration and Electrical Validation of a BIS Portable System.
    Author: Alves J, Sousa F, Correia MV.
    Journal: Stud Health Technol Inform; 2018; 249():122-129. PubMed ID: 29866967.
    Abstract:
    Heart failure (HF) affects at least 26 million people worldwide and is considered a global pandemic. Almost 1 out of 4 hospitalised patients are re-hospitalised for HF within the 30-day post-discharge period. This can be avoided if changes to the hemodynamics of the HF patient's body are early detected. Bioimpedance Spectroscopy (BIS) is a method that allows the measurement and analysis of multi-frequency body complex impedance and can be used to detect changes to the HF patient's hemodynamics. This paper presents the calibration and validation process of a low-cost portable BIS system, as well as, the study of the best Cole parameter estimation methods. The BIS system was calibrated using a three reference circuit method and a Resistance-Capacitance-Inductance (RCL) meter as calibration system. After calibration, BIS impedance measurements of validation circuits presented an average phase error of 0.76 degrees and an average magnitude relative error of 0.6% when compared with standard values. Regarding Cole parameters estimation, using the Impedance model and the Non-Linear Least Squares method for curve fitting, the relative errors were below 4% when compared with the expected values.