  • Title: CdZnTeS quantum dots based electrochemiluminescent image immunoanalysis.
    Author: Liang XL, Bao N, Luo X, Ding SN.
    Journal: Biosens Bioelectron; 2018 Oct 15; 117():145-152. PubMed ID: 29894851.
    Abstract:
    In this work, quaternary CdZnTeS quantum dots (QDs) with a particularly strong electrochemiluminescence (ECL) were synthesized as ECL signal labels. The strong ECL signals can be obtained at both cathode and anode with the ECL efficiencies of 19.78% and 1.62%, respectively. The sandwich-structured ECL immunosensors for the detection of alpha-fetoprotein (AFP) and cancer antigen 125 (CA125) were accomplished with direct ECL image analysis. Under optimal conditions, the QDs-based ECL image immunoanalysis possessed good linearity from 0.5 ng/mL to 20 ng/mL for AFP and from 20 U/mL to 500 U/mL for CA125 with the detection limit of 0.1 ng/mL and 6 U/mL, respectively (S/N = 3), and the lower detection limit obtained by photomultiplier tube were 0.1 fg/mL for AFP and 0.03 mU/mL for CA125 with the wide dynamic range from 0.5 fg/mL to 20 ng/mL and from 0.1 mU/mL to 500 U/mL, respectively (S/N = 3). Furthermore, the ECL immunoanalysis was evaluated with commercial enzyme-linked immunosorbent assay in human serum samples. The good results indicated that CdZnTeS QDs-based ECL biosensor has great potential for fast biomedical screening and multi-assays in clinical diagnosis.