Search MEDLINE/PubMed
Title: Analysis of SmartPlug Insertion-Related Complications. Author: Huang CJ, Lu CJ, Tu WH, Hou YC, Wang IJ, Hu FR, Chen WL. Journal: Eye Contact Lens; 2018 Nov; 44 Suppl 2():S333-S337. PubMed ID: 29944494. Abstract: PURPOSE: To evaluate the complications and the treatment results of SmartPlug-related complications. METHODS: Retrospective review of all patients from a single medical hospital who received SmartPlug (Medennium, Inc., Irvine, CA) insertions from October 2007 to February 2014. All patients who developed SmartPlug-related canaliculitis and pyogenic granuloma were analyzed. Lacrimal irrigation with antibiotics was performed in most patients. RESULTS: Six hundred thirty-one eyes received SmartPlug insertion and 18 eyes developed SmartPlug-related complications (14 canaliculitis and 4 pyogenic granuloma). The mean time interval from insertion to development of the complications is 3.0 (0.5-6.9) years. Sixteen eyes received lacrimal irrigation with antibiotics, and all the eyes showed improvement without recurrence. The other two eyes had recurrence of granuloma pyogenica only after surgical excision. CONCLUSIONS: SmartPlug-related complications, including canaliculitis and granuloma pyogenica, required long-term follow-up. Most of the complications can be cured by lacrimal irrigation of antibiotics.[Abstract] [Full Text] [Related] [New Search]