These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.
Pubmed for Handhelds
PUBMED FOR HANDHELDS
Search MEDLINE/PubMed
Title: High spatial resolution BOLD fMRI using simultaneous multislice excitation with echo-shifting gradient echo at 7 Tesla. Author: Su S, Lu N, Jia L, Long X, Jiang C, Zhang H, Li Y, Sun K, Xue R, Dharmakumar R, Zhang L, Liu X, Xie G. Journal: Magn Reson Imaging; 2020 Feb; 66():86-92. PubMed ID: 30172939. Abstract: We introduce an accelerated gradient echo (GRE) sequence combining simultaneous multislice excitation (SMS) with echo-shifting technique for high spatial resolution blood oxygen level dependent (BOLD) functional MRI (fMRI). The simulation was conducted to optimize scan parameters. To validate the feasibility of the proposed technique, the visual and motor task experiments were performed at 7.0 Tesla (T). The single-shot EPI sequence was also applied in comparison with the proposed technique. The simulation results showed that an optimized flip angle of 9° provided maximal BOLD contrast for our scanning scheme, allowing low power deposition and SMS acceleration factor of 5. Additionally, parallel acquisition imaging with acceleration factor of 2 was utilized, which allowed a total acceleration factor of 10 in volunteer study. The experiment results showed that geometric distortion-free BOLD images with voxel size of 1.0 × 1.0 × 2.5 mm3 were obtained. Significant brain activation was identified in both visual and motor task experiments, which were in accordance with previous investigations. The proposed technique has potential for high spatial resolution fMRI at ultra-high field because of its sufficient BOLD sensitivity as well as improved acquisition speed over conventional GRE-based techniques.[Abstract] [Full Text] [Related] [New Search]