Title: Comment on 'New photoplethysmogram indicators for improving cuffless and continuous blood pressure estimation accuracy'. Author: van Helmond N, Joseph JI. Journal: Physiol Meas; 2018 Sep 27; 39(9):098001. PubMed ID: 30183671. Abstract: OBJECTIVE: A recent study investigated the potential utility of new photoplethysmogram (PPG) indicators to improve cuffless continuous blood pressure (BP) measurement. APPROACH: In this Comment we provide additional discussion of the physiology underlying PPG- and pulse arrival time (PAT)-based BP measurement. We also discuss validation of these measurements. MAIN RESULTS: Changes in PPG features and PAT can occur independently of BP. Any study validating BP measurement based on PPG features or PAT should include a variety of calibration independent physiological challenges affecting BP. SIGNIFICANCE: PPG/PAT-based BP measurement has been investigated extensively as an alternative to cuff-based BP measurement. We feel the inherent physiological confounding in PPG/PAT-based measurement makes it unlikely that it will be of clinical value.[Abstract] [Full Text] [Related] [New Search]