Title: Performance Evaluation and Measurement Uncertainty Determination of the New Version of the Abbott ARCHITECT 25-OH Vitamin D 5P02 Assay. Author: Lim YK, Park AJ, Kweon OJ, Choi JH. Journal: Am J Clin Pathol; 2019 Jan 07; 151(2):209-216. PubMed ID: 30307462. Abstract: OBJECTIVES: We evaluated performance and measurement uncertainty of the newly released ARCHITECT 25-OH vitamin D 5P02 assay (Abbott Diagnostics, Abbott Park, IL). METHODS: In total, 300 samples were used to compare the results from ELECSYS Vitamin D Total (Roche Diagnostics, Mannheim, Germany) and ADVIA Centaur Vitamin D Total (Siemens, Tarrytown, NY). To quantify the measurement uncertainty, 25 samples and four levels of standard reference material (SRM) were measured using isotope dilution-liquid chromatography-tandem mass spectrometry according to international guidelines. RESULTS: The results of the ARCHITECT assay were equivalent to other immunoassays, but correlation coefficients were lower than the recommended criterion. SRM level 2 was considered adequate for uncertainty estimation, and the expanded measurement uncertainty of the ARCHITECT assay was 4.2%, which was superior to the other two assays. CONCLUSIONS: The restandardized ARCHITECT assay has acceptable performance in a clinical setting. However, there is still a need for further standardization of total vitamin D measurement among the automated immunoassays.[Abstract] [Full Text] [Related] [New Search]