These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.
Pubmed for Handhelds
PUBMED FOR HANDHELDS
Search MEDLINE/PubMed
Title: Liquid chromatography-tandem mass spectrometry analysis of 17-hydroxyprogesterone in dried blood spots revealed matrix effect on immunoassay. Author: Han L, Tavakoli NP, Morrissey M, Spink DC, Cao ZT. Journal: Anal Bioanal Chem; 2019 Jan; 411(2):395-402. PubMed ID: 30456606. Abstract: Immunoassays for measuring 17-hydroxyprogesterone (17-OHP) produce high rates of false positives that impact the identification of congenital adrenal hyperplasia (CAH) in neonates. A confirmatory test with high analytical specificity employing liquid chromatography-tandem mass spectrometry (LC-MS/MS) methodology is needed in newborn screening for CAH. 17-OHP and cortisol were extracted from dried blood spot (DBS) samples, resolved on a C18 column, and measured using tandem mass spectrometry. The results were compared with those determined using the AutoDELFIA immunoassay. The LC-MS/MS method had a limit of quantitation of 10.0 and 5.0 ng/mL for 17-OHP and cortisol, respectively. The method characteristics showed coefficient variation (%CV) ≤ 11.9% for both 17-OHP and cortisol, recoveries ranging from 83.1 to 101.5% for 17-OHP and from 95.1 to 102.8% for cortisol, and linearity with R2 = 0.9994 for 17-OHP and R2 = 0.9996 for cortisol, clinical sensitivity of 100.0% and a specificity of 96.4% as obtained by receiver operating characteristic analysis on 45 patient samples when 17-OHP > 39.1 ng/mL was selected as the cutoff value. Comparison between the LC-MS/MS and the AutoDELFIA immunoassay methods revealed a poor correlation for patient DBS samples (R2 = 0.6784); however, an excellent correlation was obtained for QC and proficiency test (PT) DBS samples (R2 = 0.9797). The LC-MS/MS method produced reliable results for 17-OHP and cortisol for the diagnosis of CAH. The AutoDELFIA immunoassay appears to be subject to matrix effects in the analysis for 17-OHP in DBS patient samples. The DBS samples of non-patient origin may not be suitable for assessing analytical accuracy of immunoassays.[Abstract] [Full Text] [Related] [New Search]