These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.
Pubmed for Handhelds
PUBMED FOR HANDHELDS
Search MEDLINE/PubMed
Title: Automatic determination of arsenate in drinking water by flow analysis with dual membrane-based separation. Author: Vera R, Zhang Y, Fontàs C, Almeida MIGS, Anticó E, Cattrall RW, Kolev SD. Journal: Food Chem; 2019 Jun 15; 283():232-238. PubMed ID: 30722866. Abstract: The sequential application of a polymer inclusion membrane (PIM), composed of poly(vinylidenefluoride-co-hexafluoropropylene) and the anionic extractant Aliquat 336, and a microporous polytetrafluoroethylene (PTFE) gas-permeable membrane was utilized for the first time to develop a flow analysis (FA) system, for the automatic determination of trace levels of arsenate (As(V)) in drinking water as arsine. The system incorporated a flow-through extraction cell for separation and preconcentration of arsenate and a gas-diffusion cell for the separation of arsine prior to its spectrophotometric determination based on the discoloration of a potassium permanganate solution. Under optimal conditions the FA system is characterized by a limit of detection of 3.0 μg L-1 As(V) and repeatability of 1.8% (n = 5, 25 μg L-1 As(V)) and 2.8% (n = 5, 50 μg L-1 As(V)). The newly developed FA method was successfully applied to the determination of arsenate in drinking water samples in the μg L-1 concentration range.[Abstract] [Full Text] [Related] [New Search]