These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.


PUBMED FOR HANDHELDS

Search MEDLINE/PubMed


  • Title: The accuracy evaluation of four blood glucose monitoring systems according to ISO 15197:2003 and ISO 15197:2013 criteria.
    Author: Chen H, Yao Q, Dong Y, Tang Z, Li R, Cai B, Wang R, Chen Q.
    Journal: Prim Care Diabetes; 2019 Jun; 13(3):252-258. PubMed ID: 30770203.
    Abstract:
    Self-monitoring of blood glucose is recommended to monitor glycaemic levels for the prevention and treatment of diabetes mellitus.The accuracy of blood glucose monitoring systems (BGMS) directly affects the health of the patients. In this study, we evaluated the accuracy of four different blood glucose monitoring systems, including Gold-Accu Meter, Nova StatStrip Xpress® Glucose Hospital Meter, Freestyle Optium Neo H® and Accu-Chek Performa according to ISO15197:2003 and ISO15197:2013 criteria. We found that four brands of BGMS (eight BGMS) all met the minimum requirements for system accuracy performance of ISO 15197:2003 and the accuracy criteria of ISO 15197:2013. All results of BGMS lie within zones A and B of the Consensus Error Grid (CEG). In conclusion, all BGMS met the accuracy criteria of ISO 15197:2003 and the accuracy criteria of ISO 15197:2013.
    [Abstract] [Full Text] [Related] [New Search]