  • Title: A paradigmatic case of haemolysis and pseudohyperkalemia in blood gas analysis.
    Author: Salvagno GL, Demonte D, Lippi G.
    Journal: Biochem Med (Zagreb); 2019 Feb 15; 29(1):011003. PubMed ID: 30799980.
    Abstract:
    A 51-year old male patient was admitted to the hospital with acute dyspnea and history of chronic asthma. Venous blood was drawn into a 3.0 mL heparinized syringe and delivered to the laboratory for blood gas analysis (GEM Premier 4000, Instrumentation Laboratory), which revealed high potassium value (5.2 mmol/L; reference range on whole blood, 3.5-4.5 mmol/L). This result was unexpected, so that a second venous blood sample was immediately drawn by direct venipuncture into a 3.5 mL lithium-heparin blood tube, and delivered to the laboratory for repeating potassium testing on Cobas 8000 (Roche Diagnostics). The analysis revealed normal plasma potassium (4.6 mmol/L; reference range in plasma, 3.5-5.0 mmol/L) and haemolysis index (5; 0.05 g/L). Due to suspicion of spurious haemolysis, heparinized blood was transferred from syringe into a plastic tube and centrifuged. Potassium and haemolysis index were then measured in this heparinized plasma, confirming high haemolysis index (50; 0.5 g/L) and pseudohyperkalemia (5.5 mmol/L). Investigation of this case revealed that spurious haemolysis was attributable to syringe delivery in direct ice contact for ~15 min. This case emphasizes the importance of avoiding sample transportation in ice and the need of developing point of care analysers equipped with interference indices assessment.
    [Abstract] [Full Text] [Related] [New Search]