These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.
Pubmed for Handhelds
PUBMED FOR HANDHELDS
Search MEDLINE/PubMed
Title: Clinical comparison of a new automated infrared blood culture system with the BACTEC 460 system. Author: Jungkind D, Millan J, Allen S, Dyke J, Hill E. Journal: J Clin Microbiol; 1986 Feb; 23(2):262-6. PubMed ID: 3084548. Abstract: A new blood culture instrument, the BACTEC (Johnston Laboratories, Inc., Towson, Md.) NR-660, which utilizes infrared detection of carbon dioxide from microbial metabolism, was compared with the radiometric BACTEC 460 system. There were 1,554 isolates from 18,785 paired aerobic blood cultures. Of these isolates, 1,303 were isolated from the radiometric 6B medium, and 1,259 were isolated from the NR6A medium (P = 0.06). Analysis of the data indicated no significant differences in recovery when any individual species was considered. When organisms were considered as groups, there were no significant detection differences for gram-negative bacilli, yeasts, or anaerobes. For gram-positive cocci in aerobic medium, the BACTEC 460 detected 84.3% of the total isolates, and the BACTEC NR-660 detected 79.7% (P = 0.04). There were 891 isolates from 13,983 paired anaerobic blood cultures. Of these isolates, 725 were recovered from the radiometric 7D BACTEC medium, and 723 were recovered from the NR7A BACTEC medium (P greater than 0.9). In the anaerobic media there was no significant difference in detection of any organism group, including the gram-positive cocci. When the results of the aerobic and anaerobic media were combined, there was equivalence between the two systems for the detection of gram-positive cocci (P greater than 0.2) and other organism groups. When the ability to detect septic episodes was compared, there was no significant difference for any organism group (P = 0.12). For aerobic media, the mean times for detection were 30.5 and 29.5 h for the BACTEC 460 and NR-660, respectively. For anaerobic media, the mean times for detection were 39.8 and 41.6 h for the BACTEC 460 NR-660, respectively. Compared with the BACTEC 460, the BACTEC NR-660 system had a greater ease of operation, faster test cycle, computerized data base, and equally rapid detection of positive cultures.[Abstract] [Full Text] [Related] [New Search]