These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.
Pubmed for Handhelds
PUBMED FOR HANDHELDS
Search MEDLINE/PubMed
Title: A stented bovine pericardial prosthesis in the pulmonary position. Author: Pragt H, Schoots MH, Accord RE, Arrigoni SC, Berger RM, Mariani MA, Willems TP, Ebels T, Van Melle JP. Journal: J Thorac Cardiovasc Surg; 2020 Mar; 159(3):1063-1071.e1. PubMed ID: 31400815. Abstract: BACKGROUND: Pulmonary valve replacement is very common among patients with congenital heart disease. The Carpentier Edwards Perimount valve (Edwards Lifesciences, Irvine, Calif), which was originally designed for the aortic position is among the most implanted valves. We aim to describe the follow-up of this valve in the pulmonary position. METHODS: Patients with a Perimount valve implanted between 2003 and 2013 in the University Medical Center Groningen were followed for the primary end point reintervention, defined as surgical or transcatheter valve replacement. Secondary end point was the occurrence of valve failure, defined as significant valvular regurgitation or stenosis. Explanted valves were histologically examined. RESULTS: Forty-five patients (median age at operation 27.8 years, 55.6% women) had a mean follow-up duration of 5.8 ± 3.3 years. There were 7 reinterventions (5 surgical and 2 transcatheter). Freedom from reintervention was respectively 95% ± 4% and 83% ± 8% at 5- and 10- years of follow-up. Freedom from valve failure was 75% ± 4% at 2 years, 65% ± 8% at 5 years of follow-up and 57% ± 10% at 10 years of follow-up. Morphology evaluation (n = 4) showed stiffened valves in the open position, with extensive fibrous tissue overgrowth on the leaflets and a variable proliferation of myofibroblasts. CONCLUSIONS: The Perimount valve has adequate function in the pulmonary valve position at 5 years of follow-up, although after 10 years of follow-up valve failure and reinterventions are common. Explanted valves show retraction and stiffening of the leaflets due to a fibrotic layer on both sides of the leaflet.[Abstract] [Full Text] [Related] [New Search]