These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.
Pubmed for Handhelds
PUBMED FOR HANDHELDS
Search MEDLINE/PubMed
Title: Direct antimicrobial susceptibility testing of positive blood cultures: a comparison of the Accelerate Pheno™ and VITEK® 2 systems. Author: Schneider JG, Wood JB, Smith NW, Emery CL, Davis TE, Manaloor JJ, Bocian B, Schmitt BH. Journal: Diagn Microbiol Infect Dis; 2019 Nov; 95(3):114841. PubMed ID: 31422873. Abstract: OBJECTIVES: To compare the performance and time-to-result (TTR) for antimicrobial susceptibility testing (AST) of positive blood cultures (PBC) using the Accelerate Pheno™ system (AXDX) and both a direct VITEK® 2 card inoculation workflow (DV2) and traditional FDA-approved VITEK® 2 workflow using subcultured isolates (V2). METHODS: Patient samples with monomicrobial Gram-negative rod bacteremia were tested on AXDX and DV2 in tandem and compared to V2 AST results. Categorical agreement (CA) errors were adjudicated using broth microdilution. Instrumentation times and AST TTR were compared. RESULTS: AXDX and DV2 had a CA of 93.4% and 97.4%, respectively, compared to V2. Postadjudication, AXDX, DV2, and V2 had CA of 94.7%, 95.7%, and 96.5%, respectively. Instrument run times were 6.6 h, 9.4 h, and 9.2 h, and AST TTR were 8.9 h, 12.9 h and 35.5 h, respectively. CONCLUSIONS: AXDX and DV2 ASTs are fast and reliable, which may have significant antimicrobial stewardship implications.[Abstract] [Full Text] [Related] [New Search]