Title: Ultrahigh-resolution optical coherence tomography/angiography with an economic and compact supercontinuum laser. Author: Wang TA, Chan MC, Lee HC, Lee CY, Tsai MT. Journal: Biomed Opt Express; 2019 Nov 01; 10(11):5687-5702. PubMed ID: 31799040. Abstract: In this study, a Q-switch pumped supercontinuum laser (QS-SCL) is used as a light source for in vivo imaging via ultrahigh-resolution optical coherence tomography and angiography (UHR-OCT/OCTA). For this purpose, an OCT system based on a spectral-domain detection scheme is constructed, and a spectrometer with a spectral range of 635 - 875 nm is designed. The effective full-width at half maximum of spectrum covers 150 nm, and the corresponding axial and transverse resolutions are 2 and 10 µm in air, respectively. The relative intensity noise of the QS-SCL and mode-locked SCL is quantitatively compared. Furthermore, a special processing algorithm is developed to eliminate the intrinsic noise of QS-SCL. This work demonstrates that QS-SCLs can effectively reduce the cost and size of UHR-OCT/OCTA instruments, making clinical applications feasible.[Abstract] [Full Text] [Related] [New Search]