  • Title: Evaluation of the 25-hydroxy vitamin D assay on a fully automated liquid chromatography mass spectrometry system, the Thermo Scientific Cascadion SM Clinical Analyzer with the Cascadion 25-hydroxy vitamin D assay in a routine clinical laboratory.
    Author: Benton SC, Tetteh GK, Needham SJ, Mücke J, Sheppard L, Alderson S, Ruppen C, Curti M, Redondo M, Milan AM.
    Journal: Clin Chem Lab Med; 2020 Jun 25; 58(6):1010-1017. PubMed ID: 31851610.
    Abstract:
    Background Liquid chromatography-tandem mass spectrometry (LC-MS/MS) offers advantages over immunoassay due to its increased specificity and ability to multiplex metabolites within a single run. Wide scale adoption of LC-MS/MS in routine clinical laboratories is restricted in part due to the high level of technical expertise required. The Thermo Scientific™ Cascadion™ SM Clinical Analyzer is the first fully automated, random access clinical analyser that utilises LC-MS/MS technology. We report an analytical validation of the 25-hydroxy vitamin D2 and D3 assays on the Cascadion Analyzer and an assessment of its performance within a routine clinical laboratory. Methods Analyser usability was assessed by staff with no previous experience of LC-MS/MS. An analytical validation included analysis of 154 patient samples on two different Cascadion Analyzers and a four-way method comparison of 146 patient samples on Roche and Siemens immunoassays and an in-house LC-MS/MS method. Accuracy was assessed using external quality assurance and reference materials. Seven third party IQC materials were tested on Cascadion. Results Cascadion proved easy to use by scientific and non-scientific staff. The assay passed all validation criteria. Excellent agreement was demonstrated between two different Cascadions (y = 0.97x + 3.9 nmol/L, r2 > 0.99). A method comparison demonstrated no significant difference (p > 0.05) between the Cascadion and the Roche immunoassay. A significant difference (p < 0.0001) was observed between the Cascadion and an LC-MS/MS and Siemens methods. Results obtained from EQA and reference material showed a mean bias of +3.09% and all samples were within ±10% of assigned concentrations. All third party IQC samples tested were compatible for use with Cascadion. Conclusions The Cascadion Analyzer is a fully automated LC-MS/MS system that requires no prior LC-MS/MS expertise. The vitamin D assays demonstrated excellent performance with high levels of accuracy.
    [Abstract] [Full Text] [Related] [New Search]