    Cardiac catheters assigned for reprocessing revealed only minor bacterial contamination. Random samples taken from reprocessed catheters were sterile and free from pyrogens. Among a total of 461 prospectively surveyed patients, in 16 cases a short increase of temperature up to 37.5 degrees C was observed, which appeared independent from the application of reprocessed catheters. None of the patients developed fever or symptoms of catheter-related infection.