These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.
Pubmed for Handhelds
PUBMED FOR HANDHELDS
Search MEDLINE/PubMed
Title: HPLC method transfer study for simultaneous determination of seven angiotensin II receptor blockers. Author: Marghany KA, Abdelsalam RA, Haddad GM. Journal: J Sep Sci; 2020 Apr; 43(8):1398-1405. PubMed ID: 31991050. Abstract: In this study, a sensitive high-performance liquid chromatography method was developed and validated for the simultaneous determination of seven angiotensin II receptor blockers, namely, hydrochlorothiazide, chlorthalidone, eprosartan mesylate, valsartan, losartan potassium, irbesartan, and candesartan cilexetil. Different chromatographic parameters were tested and fully optimized. Best chromatographic separation was accomplished on a reversed-phase octadecylsilyl column (250 × 4.6 mm id; 5 μm) under gradient elution using methanol/sodium phosphate monobasic buffer (0.01 M, pH 6.5) as mobile phase. The detection of target analytes was obtained at 254 nm. The pH of the buffer has been selected according to Marvin® sketch software. The proposed method was validated according to ICH guidelines and showed good precision (relative standard deviation < 1), good linearity (square of correlation coefficient ≥ 0.999), and high accuracy (between 98 and 102%) with detection limit and quantitation limit (40 and 160 ng/mL, respectively) for all the detected analytes.[Abstract] [Full Text] [Related] [New Search]