These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.


PUBMED FOR HANDHELDS

Search MEDLINE/PubMed


  • Title: Comparison of MIC Test Strip and Sensititre YeastOne with the CLSI and EUCAST Broth Microdilution Reference Methods for In Vitro Antifungal Susceptibility Testing of Cryptococcus neoformans.
    Author: Delma FZ, Al-Hatmi AMS, Buil JB, van der Lee H, Tehupeiory-Kooreman M, de Hoog GS, Meletiadis J, Verweij PE.
    Journal: Antimicrob Agents Chemother; 2020 Mar 24; 64(4):. PubMed ID: 32015025.
    Abstract:
    We compared MIC test strip (MTS) and Sensititre YeastOne (SYO) methods with EUCAST and CLSI methods for amphotericin B, 5-fluocytosine, fluconazole, voriconazole, and isavuconazole against 106 Cryptococcus neoformans isolates. The overall essential agreement between the EUCAST and CLSI methods was >72% and >94% at ±1 and ±2 dilutions, respectively. The essential agreements between SYO and EUCAST/CLSI for amphotericin B, 5-flucytosine, fluconazole, and voriconazole were >89/>93% and between MTS and EUCAST/CLSI were >57/>75%. Very major error rates were low for amphotericin B and fluconazole (<3%) and a bit higher for the other drugs (<8%).
    [Abstract] [Full Text] [Related] [New Search]