These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.
Pubmed for Handhelds
PUBMED FOR HANDHELDS
Search MEDLINE/PubMed
Title: A comparison of three column agglutination tests for red blood cell alloantibody identification. Author: Blomme S, De Maertelaere E, Verhoye E. Journal: BMC Res Notes; 2020 Mar 04; 13(1):129. PubMed ID: 32131887. Abstract: OBJECTIVE: Commercial kits of column tests for pre-transfusion testing have progressively replaced conventional tube tests in most laboratories. Aim of this study was to compare three commercial test cell panels for the identification of irregular red blood cell (RBC) alloantibodies. Overall, 44 samples with a positive indirect antiglobulin test (IAT) by routine testing were used for comparison of following panels: Ortho RESOLVE® panelC (Ortho Clinical Diagnostics (OCD), Milan, Italy), ID-DiaPanel(-P) (Bio-Rad Laboratories, CA, USA) and Identisera Diana(P) (Grifols, Barcelona, Spain). Column agglutination techniques were used, with microtubes containing either microgel (Bio-Rad/Grifols) or glass bead microparticles (Ortho). RESULTS: Alloantibody identification was possible in 38 samples, of which identical identification was shown in 33 samples by all methods. The remaining samples showed differences between certain methods, with the gel card system being superior to the glass card system for analyzing stored samples Considering that not all samples were evaluated in all three methods, the concordance rate reached 100% between Bio-Rad and Grifols, 90.5% between Bio-Rad and OCD, 86.5% between OCD and Grifols and 90.5% between all methods. Although differences in sensitivities were seen for specific antibodies, the three methods showed comparable performance for the identification of RBC alloantibodies.[Abstract] [Full Text] [Related] [New Search]