Search MEDLINE/PubMed
Title: [Indirect blood pressure determination in the horse with the Dinamap 1255 research monitor]. Author: Fritsch R, Hausmann R. Journal: Tierarztl Prax; 1988; 16(4):373-6. PubMed ID: 3222789. Abstract: In 26 horses measurings of blood pressure, using the Dinamap 1255 research monitor, were performed and compared for accuracy with blood pressure values obtained by direct measurement in the arteria carotis. It was found that oscillometrically ascertained initial blood values in standing horses have a sufficient correlation to initial blood pressure values obtained by direct measurement. But every blood pressure change is registered too late, when using the oscillometric blood pressure apparatus. This temporal delay becomes unacceptable, when a considerable increase in blood pressure occurs. A reliable oscillometric blood pressure measurement with uneven and irregular pulse waves is not possible at all.[Abstract] [Full Text] [Related] [New Search]