These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.


PUBMED FOR HANDHELDS

Search MEDLINE/PubMed


  • Title: Dielectric Properties of Glass Beads with Talc as a Reference Material for Calibration and Verification of Dielectric Methods and Devices for Measuring Soil Moisture.
    Author: Szerement J, Saito H, Furuhata K, Yagihara S, Szypłowska A, Lewandowski A, Kafarski M, Wilczek A, Majcher J, Woszczyk A, Skierucha W.
    Journal: Materials (Basel); 2020 Apr 23; 13(8):. PubMed ID: 32340125.
    Abstract:
    This paper presents dielectric measurements of talc, glass beads, and their mixtures under different moisture and salinity levels. The measurements were conducted using a prototype seven-rod probe (15 mm long central rod) connected to a single port of vector network analyzer. The samples were moistened with distilled water and KCl solutions in order to obtain six different moisture content levels. The complex dielectric permittivity was determined from vector network analyzer reflection-coefficient measurements based on the open-water-liquid calibration procedure. Next, the fitting of volumetric water content-real part of dielectric permittivity calibration curves was performed for each material at selected frequencies, and the obtained relations were compared with well-known calibration equations. Additionally, a salinity index for the tested materials was calculated. It was concluded that pure talc is not an optimal material for the calibration and verification of dielectric methods. The calibration curves obtained for glass beads and the mixtures of glass beads with talc gave results close to well-known reference calibration functions. Additionally, the addition of talc caused the data points to be less scattered. Moreover, the values of the salinity index for the tested materials were in a good agreement with literature data for sand. The obtained results indicated that glass beads with the addition of talc can be used as a reference material for the calibration and verification of dielectric methods and devices for soil moisture measurement.
    [Abstract] [Full Text] [Related] [New Search]