These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.


PUBMED FOR HANDHELDS

Search MEDLINE/PubMed


  • Title: Determination of digoxin in serum by on-line immunoadsorptive clean-up high-performance liquid chromatographic separation and fluorescence-reaction detection.
    Author: Reh E.
    Journal: J Chromatogr; 1988 Dec 09; 433():119-30. PubMed ID: 3235540.
    Abstract:
    The use of valve-switching with an anti-digoxin immunoadsorptive clean-up cartridge prior to reversed-phase elution of serum is described. Eluted digoxin is detected by reaction of the unsplit eluent stream with hydrochloric acid in a PTFE coil, followed by fluorescence detection. Routes for optimization and evaluation of the immunoadsorptive clean-up are reported, and the effects on the recovery rate examined. The detection limit of 300 pg/ml and reproducibility of less than 7% for the application of the analytical procedure to serum are reasonably good.