These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.
Pubmed for Handhelds
PUBMED FOR HANDHELDS
Search MEDLINE/PubMed
Title: Determination of cortisol in human plasma by stable isotope dilution mass spectrometry. Author: Kasuya Y, Furuta T, Hirota N. Journal: Biomed Environ Mass Spectrom; 1988 Oct; 16(1-12):309-11. PubMed ID: 3242683. Abstract: Cortisol selectively labelled with 2H at the 19-methyl and the C-1 position (cortisol-d5) was synthesized and used as an internal standard in stable isotope dilution mass spectrometry in order to determine cortisol in human plasma. A capillary gas chromatographic/mass spectrometric method provided a sensitive and reliable technique with good accuracy, precision and reproducibility without complex corrections for contributions by using cortisol-d5 as an internal standard. For calculation of plasma cortisol, peak areas were measured by selected ion monitoring on the characteristic fragment ions of the dimethoxime tri(trimethylsilyl) derivatives of cortisol and cortisol-d5 (m/z 605 and 610, respectively). The sensitivity of the gas chromatographic/mass spectrometric assay was 1.02 ng per injection, with a signal-to-noise ratio of about 6. The inter- and intra-assay coefficients of variation for plasma sample were 3.07% and 1.77%, respectively.[Abstract] [Full Text] [Related] [New Search]