These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.
Pubmed for Handhelds
PUBMED FOR HANDHELDS
Search MEDLINE/PubMed
Title: Immunoprecipitation method for CK-MB analysis re-evaluated: influence of CK-BB and macro-CK on blank activities. Author: Jacobs E, Anzalone T, Sarkozi L. Journal: Clin Chem; 1988 Mar; 34(3):585-8. PubMed ID: 3280165. Abstract: We evaluated the analytical performance of the immunoprecipitation technique for quantification of creatine kinase (CK; EC 2.7.3.2) isoenzyme MB, focusing on specimens with increased blank activities. The samples we studied had blank values that were equal to or greater than either 15 U/L or the uncorrected CK-MB value. Of 134 specimens selected, 16 and 21 contained macro CK type 1 and CK-BB, respectively. All samples containing macro CK type 1 gave negative CK-MB values, even though four contained significant quantities of CK-MB. Correcting the blank value for the fluid displacement of the second antibody suspension eliminated all but three of the negative results. However, 45 specimens, negative for CK-MB by electrophoresis, gave CK-MB values ranging from 1 to 21 U/L. Furthermore, 11 of the specimens containing macro-CK type 1 were in this group. Thus, the presence of macro-CK type 1 definitely interferes with the immunoprecipitation technique. Our results indicate the need to re-evaluate the diagnostic accuracy of the immunoprecipitation technique, especially when the un-adjusted blank activities are increased.[Abstract] [Full Text] [Related] [New Search]