These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.
Pubmed for Handhelds
PUBMED FOR HANDHELDS
Search MEDLINE/PubMed
Title: Uncoated capillary column inlets (retention gaps) in gas chromatography. Author: Grob K, Schilling B. Journal: J Chromatogr; 1987 Mar 27; 391(1):3-18. PubMed ID: 3294873. Abstract: Uncoated but deactivated pre-columns have become a widely used tool in capillary gas chromatography (GC), serving strongly differing purposes. Pre-columns are often used as guard columns, reducing the effects of involatile sample by-products on chromatographic performance and rendering exchange of contaminated column inlets simple. Wide-bore pre-columns facilitate introduction of the syringe needle and open the way for a relatively robust on-column autosampler. Other pre-columns are used for re-concentrating solute bands that are broadened due to the flow of sample liquid in the column inlet (retention gap). Long pre-columns allow on-column injection of large sample volumes (e.g., 50-80 microliter when a 15 m X 0.32 mm I.D. pre-column is used). The background of the various uses of pre-columns is discussed, concluding with an evaluation of different deactivation methods for the internal wall of the pre-columns. Critical parameters are inertness, wettability and retention power. Press-fit connections are recommended for coupling pre-columns to the coated columns.[Abstract] [Full Text] [Related] [New Search]