Title: [Accidental entrapment of a pulmonary artery catheter during heart transplantation]. Author: Bleiler S, Holzamer A. Journal: Anaesthesist; 2020 Nov; 69(11):821-824. PubMed ID: 32964247. Abstract: Inadvertent fixation of a Swan-Ganz pulmonary artery catheter by a suture is a rare complication. We present a case report of a patient undergoing heart transplantation in whom the catheter was fixed by a suture at the upper venous cannulation site for the cardiopulmonary bypass. After diagnostics with fluoroscopy the chest was reopened and the suture was released. The catheter could then be removed. The patient did not suffer from any consequential damage. Die akzidentelle Fixierung eines Pulmonalarterienkatheters durch eine Naht ist eine seltene Komplikation. Dieser Fall handelt von einem Patienten zur Herztransplantation, bei dem der Katheter mittels Naht an der oberen venösen Kanülierungsstelle der Herz-Lungen-Maschine fixiert wurde. Nach Diagnostik mittels Fluoroskopie erfolgten die Resternotomie und die Lösung der Naht, woraufhin sich der Katheter entfernen ließ. Der Patient trug keine Folgeschäden davon.[Abstract] [Full Text] [Related] [New Search]