These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.


PUBMED FOR HANDHELDS

Search MEDLINE/PubMed


  • Title: Early experience with the Micro Plug Set for preterm patent ductus arteriosus closure.
    Author: Heyden CM, El-Said HG, Moore JW, Guyon PW, Katheria AC, Ratnayaka K.
    Journal: Catheter Cardiovasc Interv; 2020 Dec; 96(7):1439-1444. PubMed ID: 32979038.
    Abstract:
    OBJECTIVES: We intend to describe early experience using a new, commercially available Micro Plug Set for preterm neonate and infant transcatheter patent ductus arteriosus (PDA) occlusion. BACKGROUND: Transcatheter PDA occlusion in premature neonates and small infants is safe and effective. The procedure is early in its evolution. METHODS: Procedural and short-term outcomes of preterm neonates and infants undergoing transcatheter PDA occlusion with a new, commercially available device were reviewed. RESULTS: Eight preterm neonates and infants born at median 27 weeks gestation (23-36 weeks) underwent transcatheter PDA device closure with the Micro Plug Set. The device is short (2.5 mm) with a range of diameters (3, 4, 5, 6 mm) and delivered through a microcatheter. Procedures were performed at median 41 days of age (12-88 days) and at 1690 g (760-3,310 g). Transvenous PDA device occlusion was performed with fluoroscopic and echocardiography guidance. All procedures were successful with complete PDA occlusion. There were no procedural or short-term adverse events. CONCLUSIONS: Preterm neonate and infant transcatheter PDA device closure with a new, commercially available short and microcatheter delivered device (Micro Plug Set) was safe and effective in a small, early series of patients.
    [Abstract] [Full Text] [Related] [New Search]