These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.


PUBMED FOR HANDHELDS

Search MEDLINE/PubMed


  • Title: Corrosion of ferromagnetic alloys used for magnetic retention of overdentures.
    Author: Vrijhoef MM, Mezger PR, Van der Zel JM, Greener EH.
    Journal: J Dent Res; 1987 Sep; 66(9):1456-9. PubMed ID: 3305638.
    Abstract:
    Three ferromagnetic casting alloys intended for use with magnetically retained overdentures (compositions in wt%: #1, Pd 50, Co 47, Ga 2, Pt 1; #2, Pd 60, Co 37, Ga 2, Pt 2; #3, Pd 49, Co 46, Ga 2, Pt 2) were investigated. A 0.9% saline solution and an artificial saliva (Meyer) were used for evaluation of the electrochemical corrosion behavior of these alloys, utilizing standard potentiodynamic techniques. All alloys investigated possessed a good corrosion resistance in the potential range of the oral environment (from -100 to 300 mV versus SCE). In addition, alloys #1 and 3 appeared to be susceptible to pitting above +300 mV (SCE), while #2 was resistant to pitting and was found to passivate in 0.9% saline solution.
    [Abstract] [Full Text] [Related] [New Search]