These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.
Pubmed for Handhelds
PUBMED FOR HANDHELDS
Search MEDLINE/PubMed
Title: Determination of phenobarbital and phenytoin in serum by a mechanized enzyme immunoassay (EMIT) in comparison with a gas-liquid chromatographic method. Author: Oellerich M, Külpmann WR, Haeckel R, Heyer R. Journal: J Clin Chem Clin Biochem; 1977 Jun; 15(6):353-8. PubMed ID: 330795. Abstract: The antiepileptic drugs phenobarbital and phenytoin were determined in serum by enzyme immunoassay (Emit, Syva Corp.) and gas-liquid chromatography. The Emit assays were mechanized by the use of an Eppendorf analyzer 5010. The precision of the Emit system was sufficient (coefficient of variation within series 6-13% and from day to day 8-15% with various calibrators and control sera). Moreover the Emit method is rapid, specific and easy to perform. The procedure requires only 10 microliter of serum per determination. A disadvantage however is the high cost of the reagents. A comparison of the results obtained by Emit and gas-liquid chromatography in a series of about 50 patients showed a good correlation between both methods (correlation coefficient r = 0.968 for phenobarbital and 0.978 for phenytoin).[Abstract] [Full Text] [Related] [New Search]