These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.


PUBMED FOR HANDHELDS

Search MEDLINE/PubMed


  • Title: Long period grating in double cladding fiber coated with graphene oxide as high-performance optical platform for biosensing.
    Author: Esposito F, Sansone L, Srivastava A, Baldini F, Campopiano S, Chiavaioli F, Giordano M, Giannetti A, Iadicicco A.
    Journal: Biosens Bioelectron; 2021 Jan 15; 172():112747. PubMed ID: 33129073.
    Abstract:
    In this work, the development and testing of a novel fiber-optic based label-free biosensor is presented, whose performance were verified through the detection of C-reactive protein (CRP) in serum. The device is based on a long period grating fabricated in a double cladding fiber with a W-shaped refractive index (RI) profile. As a result, the working point of the device was tuned to the mode transition region by chemical etching of the outer fiber cladding, obtaining a significant enhancement of the RI sensitivity and an excellent visibility of the grating resonances due to the mode transition in an all-silica structure. The fiber transducer was coated with a nanometric thin layer of graphene oxide in order to provide functional groups for the covalent immobilization of the biological recognition element. A very low limit of detection of about 0.15 ng/mL was obtained during the detection of CRP in serum, and a large working range (1 ng/mL - 100 μg/mL) of clinical relevance has been also achieved.
    [Abstract] [Full Text] [Related] [New Search]