Search MEDLINE/PubMed
Title: [Research of Methods to Reduce Alarm Fatigue of Monitoring System]. Author: Liu M, Sun Z, Ye W, Liu S, He X, Wang C, Li Y. Journal: Zhongguo Yi Liao Qi Xie Za Zhi; 2020 Dec 08; 44(6):481-486. PubMed ID: 33314853. Abstract: OBJECTIVE: In order to solve alarm fatigue, the algorithm optimization strategies were researched to reduce false and worthless alarms. METHODS: A four-lead arrhythmia analysis algorithm, a multiparameter fusion analysis algorithm, an intelligent threshold reminder, a refractory period delay technique were proposed and tested with collected 28 679 alarms in multi-center study. RESULTS: The sampling survey indicate that the 80.8% of arrhythmia false alarms were reduced by the four-lead analysis, the 55.9% of arrhythmia and pulse false alarms were reduced by the multi-parameter fusion analysis, the 28.0% and 29.8% of clinical worthless alarms were reduced by the intelligent threshold and refractory period delay techniques respectively. Finally, the total quantity of alarms decreased to 12 724. CONCLUSIONS: To increase the dimensionality of parametric analysis and control the alarm limits and delay time are conducive to reduce alarm fatigue in intensive care units.[Abstract] [Full Text] [Related] [New Search]