These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.


PUBMED FOR HANDHELDS

Search MEDLINE/PubMed


  • Title: Infrared tympanic thermometer: evaluation of a new clinical thermometer.
    Author: Shinozaki T, Deane R, Perkins FM.
    Journal: Crit Care Med; 1988 Feb; 16(2):148-50. PubMed ID: 3342626.
    Abstract:
    We have evaluated a new clinical thermometer for accuracy. It determines temperature by measuring infrared radiation given off by a warm object. It was found to be accurate both in vivo and in vitro over the temperature range of 34.0 degrees to 39.5 degrees C. The in vivo assessment was made in patients in the ICU after their return from open heart surgery. The infrared tympanic thermometer tracked the core temperature (as measured by the thermistor tip of the pulmonary artery catheter) closely, with a correlation coefficient of 0.98, and took less than 2 sec to measure. We found this new thermometer to be accurate and, in our opinion, easily usable in the clinical situation.
    [Abstract] [Full Text] [Related] [New Search]