Search MEDLINE/PubMed
Title: Liquid chromatographic determination of flucytosine in capsules: collaborative study. Author: Shostak D, Klein C. Journal: J Assoc Off Anal Chem; 1988; 71(1):33-5. PubMed ID: 3360702. Abstract: A liquid chromatographic method for the determination of flucytosine in capsules was collaboratively studied by 7 laboratories. The method uses a C18 reverse phase column, water-methanol-acetic acid mobile phase containing 1-octanesulfonic acid sodium salt, p-aminobenzoic acid as internal standard, and photometric detection at 285 nm. The mean recovery value (+/- SD) of flucytosine from a synthetic formulation representing capsules was 99.2 +/- 1.72% (CV = 1.73%). Composited samples of 250 and 500 mg commercial capsules gave assay values of (mean +/- SD) 103.17 +/- 2.21 and 99.29 +/- 1.29% of declared, respectively. CV values were 2.15 and 1.30%. Reproducibility and repeatability CVs were 2.19 and 1.50%, respectively, for the 250 mg capsules, and 1.34 and 0.63%, respectively, for the 500 mg capsules. The method has been adopted official first action.[Abstract] [Full Text] [Related] [New Search]