These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.


PUBMED FOR HANDHELDS

Search MEDLINE/PubMed


  • Title: Revised calibration of the Reflotron cholesterol assay evaluated.
    Author: Boerma GJ, van Gorp I, Liem TL, Leijnse B, Bellm J, Carstensen CA.
    Journal: Clin Chem; 1988 Jun; 34(6):1124-7. PubMed ID: 3378332.
    Abstract:
    We evaluated the Boehringer Mannheim (B.M.) Reflotron Total Cholesterol "dry-chemistry" method after its recalibration in 1987. Reports in the literature up to 1986-1987 of a negative bias (up to -10%) in the method prompted a revision of the factory-set calibration of the Reflotron. For this, B.M. prepared a new set of calibrators with 12 different concentrations of cholesterol. We checked in two ways whether accuracy had been achieved: (a) The values assigned to the calibrators by B.M. were checked with the manual Abell-Kendall Reference Method (MAK) performed in an official Reference Center. These were shown to be correct. (b) Concurrently, a direct comparison was made by analyzing 200 fresh samples of human serum. Reflotron cholesterol values obtained for these samples proved to be accurate, meeting the current World Health Organization/Centers for Disease Control criterion of maximum bias less than or equal to 5%. Orthogonal regression analysis yielded the following correlation: Reflotron = 0.985 MAK + 0.238 mmol/L (y = ax + b). Reflotron mean = 6.26 mmol/L; MAK mean = 6.09 mmol/L. SDa = 0.015 mmol/L; SDb = 0.120 mmol/L, and r = 0.989.
    [Abstract] [Full Text] [Related] [New Search]