These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.


PUBMED FOR HANDHELDS

Search MEDLINE/PubMed


  • Title: Bias current modifies the selectivity of liquid membrane ion-selective microelectrodes.
    Author: Coles JA.
    Journal: Pflugers Arch; 1988 Mar; 411(3):339-44. PubMed ID: 3380648.
    Abstract:
    A negative bias potential of up to -80 mV applied to the back of a liquid membrane ion-selective microelectrode containing classical "K+" ion-exchanger was found to make it more selective for millimolar concentrations of K+ over micromolar concentrations of choline, tetramethylammonium, tetraethylammonium and 5-hydroxytryptamine. Conversely, positive bias potential increased severalfold the sensitivity to micromolar concentrations of these ions while decreasing the sensitivity to K+. An increase in response amplitude for millimolar changes of ion concentration was also observed in neutral carrier electrodes for Na+, K+ and Ca2+ with negative bias potential. The various ions caused the resistances of the electrodes to change; these resistance changes contributed to the changes in response amplitude, but there were additional, unexplained, factors. The phenomenon was used to test if the signal from a K+ ion-exchanger microelectrode in extracellular space in bee retina was contaminated by substances other than K+.
    [Abstract] [Full Text] [Related] [New Search]