These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.


PUBMED FOR HANDHELDS

Search MEDLINE/PubMed


  • Title: Instant photographic refractometry in children.
    Author: Wanger P, Waern G.
    Journal: Acta Ophthalmol (Copenh); 1988 Apr; 66(2):165-9. PubMed ID: 3389090.
    Abstract:
    A simple photographic method for detection and measurement of refractive errors in children, using a specially designed camera and electronic flash unit and 'instant' (Polaroid) film, was tested on 64 children, aged 3 to 8 years, and compared with the results from retinoscopy. The refractive errors ranged from -6.5 to +6.5 D. A light reflex in the pupil was observed on the photos of all children with hyperopia greater than = 0.5 and myopia greater than = 2.0 D. A curvilinear relationship was observed (in the range -2 to -4 and +0.5 to +3 D) between the width of the light reflex in the pupil, easily measured on the photos and the degree of refractive error, determined by retinoscopy. The degree of ametropia could be estimated with reasonable accuracy (+/- 0.5 D) in the interval from -2 to -4 D and +0.5 to +3 D. The method can demonstrate the type and, to a limited extent, also the degree of refraction anomaly and may be of value in screening for refractive errors in children.
    [Abstract] [Full Text] [Related] [New Search]