These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.
Pubmed for Handhelds
PUBMED FOR HANDHELDS
Search MEDLINE/PubMed
Title: Environmental temperature control in very low birth weight infants (less than 1000 grams) cared for in double-walled incubators. Author: Chessex P, Blouet S, Vaucher J. Journal: J Pediatr; 1988 Aug; 113(2):373-80. PubMed ID: 3397804. Abstract: To evaluate the effect of fluctuations in environment and body temperatures on preterm infants, we recorded these variables in very immature newborn infants (birth weight less than 1000 gm) cared for in double-walled incubators (Air-Shields model C-100 and Ohio model IC). Both incubators maintained environmental temperatures corresponding overall to the set point, despite incubator openings. Under skin temperature servocontrol, however, environmental temperature fluctuations were greater than 2 degrees C even in strictly controlled conditions. The pattern of incubator temperature fluctuations depended on the set point rather than on the type of incubator (conventionally heated or heated by warm air blown between the double walls). The long-term clinical significance of the incubator temperature variability remains to be determined; the choice between air and skin servocontrolling should depend in part on the need for environmental stability.[Abstract] [Full Text] [Related] [New Search]