These tools will no longer be maintained as of December 31, 2024. Archived website can be found here. PubMed4Hh GitHub repository can be found here. Contact NLM Customer Service if you have questions.
Pubmed for Handhelds
PUBMED FOR HANDHELDS
Search MEDLINE/PubMed
Title: A 4-μW Analog Front End Achieving 2.4 NEF for Long-Term ECG Monitoring. Author: Yang W, Jiang H, Yin Y, Wang Z. Journal: IEEE Trans Biomed Circuits Syst; 2021 Aug; 15(4):655-665. PubMed ID: 34043513. Abstract: An ultra-low-power low-noise analog front end (AFE) is presented in this work, aiming for long-term ECG with clear P-waves for clinical diagnose. The chopper amplifier with passive noise filter and PWM based offset cancellation results in an input-reference noise of 0.39 μVrms, which shows 3.7X noise improvements among the state-of-the-art designs. With the digital offset cancellation by the pulse-width modulation wave, the AFE achieves a low input-referred dc offset of 0.4 μV among 9 tested chips and a low drift under 30 μV. A dynamic scale ADC with low-power comparator strategy prevents the instability and signal-loss, achieving an SFDR of 71.6 dB. The proposed AFE achieves a noise-efficient-factor (NEF) of 2.4 with a power consumption of 4 μW. The fabricated chip is demonstrated in a miniature prototype for long-term ECG monitoring application, presenting a clear ECG waveform with visible P-wave. The simultaneously ECG recording with a medical grade 12-lead ECG Holter shows the effective acquisition of the prototype, proofing the better noise performance.[Abstract] [Full Text] [Related] [New Search]